Description
Specifications:
- Plate types: 6 – 1536-well plates
- Wavelength selection: Double excitation and double emission monochromators
- Excitation wavelength range: 200 – 1000 nm
- Emission wavelength range: 270 – 840 nm
- Excitation/emission bandwidth: 5 nm and 12 nm/12 nm
- Light source: Xenon flash lamp
- Sensitivity/dynamic range:
-
- Fluorescence intensity, top reading: < 0.4 fmol fluorescein/well, > 6 decades, 384-well plate
- Fluorescence intensity, bottom reading: < 4 fmol fluorescein/well, > 5.5 decades, 384-well plate
- Time-resolved fluorescence, top reading: < 120 amol Europium/well, > 6 decades, 384-well plate
-
- Wavelength selection: All wavelengths, filters and double monochromators
- Wavelength range: 360 – 670 nm, spectral scanning 270 – 840 nm
- Sensitivity/dynamic range: < 7 amol ATP/well, > 7 decades, flash ATP reaction, 384-well plate
- Plate types 6 – 384-well plates
- Wavelength selection: Double monochromators
- Wavelength range: 200 – 1000 nm
- Bandwidth: 5 nm
- Light source: Xenon flash lamp
- Dispenser up to 3, automatic dispensing position control
- Dispensing volume: 1 – 10 000 µl, with 1 µl increments (1 ml syringe)
- Precision: 5 – 19 μl < 5%, 20 – 10 000 μl < 2% (1 ml syringe, 0.40 mm tip)
- Dispensing speed: 30 s, 96-well plate 80 s, 384-well plate (5 µl/well, 1 ml syringe, 0.40 mm tip)
- Incubator: From ambient + 4°C to 45°C at ambient 25°C
- General Features: Measurement modes Endpoint, kinetic, spectral scanning with all detection technologies
- Measurement speed: 96-well plate in 15 s, 384-well plate in 45 s, and 1536-well plate in 135 s (minimum kinetic interval time from A1 back to A1)
- Overall dimensions: 540 mm (W) x 580 mm (D) x 500 mm (H)
[21.3” (W) x 22.8” (D) x 19.7” (H)]