Description
Multi-purpose systems providing fast, efficient cell fusion in hybridoma production, hybrid cell formation and nuclear transfer applications, as well as square wave electroporation for transfecting genes and other molecules into mammalian cell lines. Meet your exact needs with preconfigured systems and a choice of basic or advanced systems.
Specifications:
Firmware Specifications;
Electrofusion Quick Start Menu: Up to 3 protocol steps, ordered AC-DC-AC
Electrofusion User Defined Protocols: Up to 19 AC protocol steps and up to 1 DC step, organized in any order
Electrofusion Preset Protocols: Over 20 Preset Protocols for electrofusion applications
Electroporation Quick State Menu: Yes
Electroporation User Defined Protocols: Yes
Square Wave Pulse, DC;
Voltage Range:
LV Mode: 5 to 500 V in 1 V steps
HV Mode: 505 to 3000 V in 1 V steps
Voltage Accuracy:
LV Mode: 5%
HV Mode: 5%
Pulse Length Range:
LV Mode: 10 to 999 µs in 1 µ steps
LV Mode: 1 to 999 ms in 1 ms steps
HV Mode: 10 to 600 µs in 1 µs steps
Multiple Pulsing
LV Mode: 0 to 100 pulses per sample
HV Mode: 0 to 100 pulses per sample
Pulse Interval: 0.1 s to 10 s
Alignment (Pre-Pulse), AC;
Frequency: 0.2 to 2 MHz in 0.1 MHz steps
Voltage: 5 to 75 V in 5 V steps
Duration: 0 to 99 s in 1 s steps
Wave Shape: Sine wave
Post-Fusion,AC;
Frequency: 0.2 to 2 MHz in 0.1 MHz steps
Voltage: 5 to 75 V in 5 V steps
Duration: 5 to 75 V in 5 V steps
Wave Shape: Sine wave
Other Specifications;
Charging Time: LV = 7 s, HV = 7 s
DC Monitoring:
Protocol Name, Steps Executed,
Time/Date, Error Conditions, Pulse
Voltage, Pulse Current,
Pulse Width, Pulse #, Droop %,
Sample Voltage, Sample Current and
Sample Resistance
Display: 7-inch full color display
Controls: Touch Screen
Programmability: Stores over 1000 protocols
DC Arc Control: Yes
Pre-Pulse Sample Load Check: Ves (20 VDC)
Computer (Optional): PC running Microsoft Windows XP, 7, 8.1 or 10
Communications: USB
Data File Format: Comma Separated Values (CSV)
Remote Operation: Yes, using Foot Pedal control
Accessory Connector Type: Coaxial
Power Ratings: 70 W idle and 900 W pulsing
Input Voltage Ratings: 100 to 240 VAC, 50/60 Hz
Dimensions (H x W x D): 13 x 12 x 13 in
Weight: 22 lb
Atmospheric Specifications
Operating Temperature: 4°C to 40°C (40°F to 104°F)
Storage Temperature: -10°C to 70°C (14°F to 158°F)
Operating Humidity: 20% to 80% RH, non-condensing
Storage Humidity: 20% to 80% RH, non-condensing
Mode of Operation: Continuous
Classification: Class I
Pollution: IP2X
Installation: Category II