Description
The Gene Pulser Xcell System is a flexible, modular electroporation system for transfecting every cell type from primary, suspension, and difficult-to-transfect cells, including T cells, to bacteria and fungi. The system is composed of a main unit, two accessory modules, the capacitance extender (CE module) and the pulse controller (PC module), and a ShockPod cuvette chamber.
Specifications:
Waveform Output: Exponential or square
Voltage Range:
Main Unit & Microbial System: 200–3,000 V
Eukaryotic System: 10–3,000 V
Capacitance:
10–500 V: 25–3,275 µF (25 µF increments)
500–3,000 V: 10, 25, 50 µF
Parallel Resistance: 50–1,000 Ω (50 Ω increments) + infinity (not in eukaryotic system)
Sample Resistance:
20 Ω minimum at 10–2,500 V (or 200–2,500 V in microbial system)
600 Ω minimum at 2,500–3,000 V
Square-Wave Pulse Timing:
10–500 V: 0.05–10 ms (0.05 ms increments), 10–100 ms (1 ms increments), 1–10 pulses, 0.1–10 s interval
500–3,000 V: 0.05–5 ms (0.05 ms increments), 1–2 pulses, 5 s minimum interval
Power Input: 100–120 VAC or 220–240 VAC, 50/60 Hz
Power Consumption: Max. 240 W (during charging)
Operating Conditions: 0–35°C; 0–95% RH (noncondensing)
Regulatory Compliance:
Safety: EN 61010
EMC: EN 61326 Class A
Dimensions:
Main Unit: 31 x 30 x 14 cm
CE Module: 31 x 28 x 9 cm
PC Module: 31 x 28 x 5 cm
Weight:
Main Unit: 6.6 kg
CE Module: 3.1 kg
PC Module: 1.9 kg